The EOS Power MWLP225-1003-II is an open-frame AC/DC converter designed for medical applications. It operates with a universal input voltage range of 85-264 VAC and provides up to 120W of power via natural convection cooling, or 225W with forced air cooling.
Key electrical specifications include a 93% efficiency, power factor exceeding 0.95, and a hold-up time of 10ms at 225W. The converter features over-current, over-voltage, and short-circuit protection. It meets IEC60601-1-2 standards for medical device electromagnetic compatibility.
With dimensions of 4.00" L x 2.00" W x 1.00" H, this chassis-mount unit is suitable for integration into various systems. The operating temperature range is -40°C to 70°C, and it boasts a Mean Time Between Failures (MTBF) of 3.37 million hours.
This power supply is ideal for medical equipment requiring reliable and safe power conversion. Its robust design and compliance with medical safety standards make it a suitable choice for demanding applications.
The output power is 120W with natural convection cooling and up to 225W with forced air cooling.
The input voltage range is 85-264 VAC.
The efficiency is up to 94% depending on the output voltage.
The operating temperature range is -40°C to 70°C.
Yes, it meets Medical (BF) Safety Approvals and standard IEC60601-1-2: 2014 (4th Edition).
The MTBF is 3.37 million hours, calculated per Telcordia-SR332-issue 3.
The dimensions are 4.00 inches long by 2.00 inches wide by 1.00 inch high.