| Category | |
| Manufacturer Name | |
| Efficiency | |
| Medical Grade | |
| Mounting Type | |
| Operating Temperature | |
| Output Current | |
| Output Power | |
| Output Voltage | |
| Protection | |
| Relative Humidity | |
| Ripple | |
| Storage Temperature | |
| Voltage | |
| Weight |
The Autec Power Systems WM-M020A-059-C-SUSB is a medical-grade AC/DC wall mount power adapter designed for reliable power delivery in medical equipment applications. This adapter offers a single output of 5.9V at 3A, providing up to 17.7W of power.
It features a universal 100-240Vac input range and meets Efficiency Level VI standards. Safety compliance includes IEC/EN/ANSI/AAMI ES60601-1 (edition 3.1) and EMC compliance with IEC60601-1-2 (edition 4.0). The adapter also adheres to EMI standards EN55011 Class B and harmonic current standards EN61000-3-2, 3.
Key electrical specifications include regulated output with low ripple noise, load regulation of ±5% typical, and a transient response of 0.5mS for a 50% load change. Protection features include internal primary current fuse, inrush limiting, short circuit, over voltage, and over current protection. The leakage current is less than 100µA.
Designed for wall mounting, this adapter comes with various plug type options (US, EU, UK). It operates within a temperature range of 0 to +40°C and has a calculated MTBF of 300,000 hours at 25°C. The series is noted as 'Not Recommended for New Design', with WM-M024A recommended as a replacement.
The WM-M020A-059-C-SUSB provides an output voltage of 5.9V at a current of 3A, delivering 17.7W of power.
This adapter supports a universal input voltage range of 100-240Vac.
The adapter complies with IEC/EN/ANSI/AAMI ES60601-1 (edition 3.1) and EMC compliance with IEC60601-1-2: 2014 (edition 4.0).
The adapter meets Efficiency Level VI.
The operating temperature range is from 0 to +40°C.
It includes internal primary current fuse, inrush limiting, short circuit protection, over voltage protection, and over current protection.
No, this part is not recommended for new design. The recommended replacement is the WM-M024A.