| Category | |
| Manufacturer Name | |
| Applications | |
| Approval Agency | |
| Cord Length | |
| Current | |
| Efficiency | |
| Form | |
| Height | |
| Input Connector | |
| Input Type | |
| Length | |
| No Load Power Consumption | |
| Operating Temperature | |
| Output Connector | |
| Polarization | |
| Power (Watts) | |
| Region Utilized | |
| Standard Number | |
| Storage Temperature | |
| Voltage | |
| Width |
The Cincon TR220MA360-01E13 VI is a 220 Watt AC/DC desktop adapter designed for medical applications. It features a universal input range of 80-264Vac and provides a stable 36V output at 6.11A.
This adapter boasts high efficiency up to 93% and a low no-load power consumption of less than 150mW. It meets stringent medical safety standards, including IEC/EN/UL 60601-1 (2 MOPP) and IEC/EN 60601-1-11 for home healthcare environments. The unit also complies with EN55011 and CISPR/FCC Class B for electromagnetic interference.
Key electrical specifications include a typical line regulation of ±1% and load regulation of ±2%. The operating temperature range is -30°C to 70°C. The adapter has a Class I form factor and comes with an IEC 320-C14 input connector and a 1.80m cord with a 2.1mm x 5.5mm DC output plug.
It includes continuous short circuit protection and over voltage protection. The dimensions are 180.0mm x 75.0mm x 44.5mm. This adapter is suitable for international use.
100 to 240 V AC.
70 degrees Celsius.
36 V DC at 6.11 A.
4000 V AC between input and output.
93 percent at 230 V AC and 75 percent full load.
180.0 mm length, 75.0 mm width, and 44.5 mm height.