The EOS Power MWLP350-1306-II is a versatile open-frame AC/DC power supply unit designed for demanding applications, particularly in the medical field. It offers a robust output power of up to 350W with forced air cooling or 200W with natural convection, operating from a universal input voltage range of 90-264 VAC.
Key electrical specifications include high efficiency up to 94%, a power factor exceeding 0.95 at full load, and comprehensive protection features such as over-current, over-voltage, and short-circuit protection in hiccup mode with auto-recovery. The unit also provides a 12V/0.5A fan output and includes a thermal shut-down feature.
Designed for reliability, the MWLP350-1306-II boasts a Mean Time Between Failures (MTBF) of 2.56 million hours (Telcordia SR-332 Issue 3). It meets stringent medical safety standards, including IEC60601-1-2:2014 (4th Edition) and carries BF type isolation ratings. The operating temperature range is -40°C to 70°C, with startup guaranteed even at lower temperatures.
This power supply is chassis mountable and has dimensions of 5.00" L x 3.00" W x 1.00" H. Its open-frame construction and medical certifications make it suitable for integration into various medical devices and other industrial equipment requiring a reliable and efficient power source.
The MWLP350-1306-II can deliver up to 350W with forced air cooling or 200W with natural convection cooling.
The input voltage range is 90-264 VAC.
The operating temperature range is -40°C to 70°C.
Yes, it meets standard IEC60601-1-2:2014 (4th Edition) and has Medical (BF) safety approvals.
It is a chassis mount type.
The typical MTBF is 2.56 million hours, based on Telcordia SR-332 issue 3.