The EOS Power MULP40-1312 is an open-frame AC/DC converter designed for medical applications, delivering 40W of power with a single 12V output. It accepts a universal input voltage range of 85-264 VAC and operates at efficiencies up to 85% typical.
This converter features a compact form factor measuring 3.00 x 2.00 x 0.75 inches, suitable for chassis mounting. It is approved to EN60601 3rd Edition and meets standard IEC60601-1-2:2014 (4th Edition), ensuring suitability for medical equipment. The operating temperature range is -40°C to 70°C, with dual fusing for added safety.
Key electrical specifications include line regulation of +/-0.5%, load regulation of +/-1%, and a hold-up time of 60 ms at 230 VAC. Over-current protection is set at >110%, and over-voltage protection ranges from 110 to 140%. The MTBF is rated at 2 million hours per Telcordia-SR332-issue 3.
Additional features include standby output and universal input capabilities. The isolation voltage is rated at 4000 VAC (input to output), 1500 VAC (input to ground), and 1500 VAC (output to ground for type BF).
The input voltage range is 85-264 VAC universal, with 47-63 Hz frequency.
The output is 12V DC at 3.33A, providing 40W total power.
It is approved to EN60601 3rd Edition and meets IEC60601-1-2 (4th Edition) EMC standards.
The operating temperature range is -40°C to +70°C with derating.
It includes over-current protection (>110%), over-voltage protection (110-140%), and short circuit protection (hiccup mode).
Typical efficiency is 85%.
The open frame converter measures 3.00 x 2.00 x 0.75 inches (76.2 x 50.8 x 19.1 mm) and is chassis mount.