The Inventus Power MSM6324 is a 60 Watt medical-grade AC/DC power supply, offering a 24V output and a 2.5A current rating. It operates with a wide input voltage range of 100-240 VAC and a frequency of 47-63 Hz, making it suitable for global applications. This open-frame converter boasts a high efficiency of 80% minimum at full load and 115 VAC.
Designed for reliability, the MSM6324 includes built-in over-voltage, over-current, and short-circuit protection. It has a compact form factor measuring 4.00 x 2.00 x 0.92 inches, suitable for chassis mounting. The operating temperature range is 0°C to 50°C, with forced air required for full load operation. It meets FCC class B and EN 55011 standards for EMI/EMC, and complies with EN60601-1 for medical safety.
Key electrical specifications include typical line regulation of +/- 0.5% and load regulation of +/- 2%. The hold-up time is a minimum of 11 ms at full load and 115 VAC. Transient response is 0.5 ms for a 50% load change. The power supply is CE marked and carries UR/cUR UL2601-1 approvals, underscoring its suitability for medical equipment.
This power supply is designed for applications requiring a stable, regulated DC output from an AC source, particularly in the medical field where safety and reliability are paramount. Its open-frame construction and chassis mount type facilitate integration into larger systems.
The MSM6324 provides a nominal 24V output at 2.5A.
The input voltage range is 100-240 VAC at 47-63 Hz.
The maximum power is 60W with at least 0.5 CFM forced air cooling, or 50W with convection cooling.
The MSM6324 includes built-in over voltage protection, over current protection, and short circuit protection.
The converter measures 101.6mm x 50.8mm x 23.3mm and is designed for chassis mount.
The operating temperature range is 0°C to 50°C with forced air cooling for full load.
It meets FCC Class B, EN 55011, and EMC per EN60601-1-2.