The SL Power Electronics MENB1060A1802F01 is a 61W AC-DC external power adapter designed for medical applications. It operates with a universal input voltage range of 100 to 240 VAC and provides a single, regulated 18V output.
This adapter meets Level V efficiency standards and features a low no-load power consumption of less than 0.5W. It is built with an impact-resistant polycarbonate enclosure and includes safety approvals to EN/IEC/UL60601-1, satisfying 2 MOPP requirements. The unit is designed for environments where AC mains power may be unstable.
Key electrical specifications include a 1.5A input current at 100 VAC, 18 mS hold-up time at 115 VAC, and ripple and noise within 1% pk-pk. Protection features include overload and short circuit protection in hiccup mode. The adapter has a Mean Time Between Failures (MTBF) of over 100,000 hours and comes with a 3-year warranty.
The MENB1060A1802F01 has a desktop form factor (Class I) with dimensions of 108.0mm x 67.7mm x 32.5mm. It utilizes an IEC 320-C14 input connector and a 2.1mm I.D. x 5.5mm O.D. x 9.5mm barrel plug output connector with positive center polarization. The operating temperature range is 0°C to 40°C.
The output voltage is 18V.
The maximum output power is 61W.
The input voltage range is 100 to 240 VAC.
The operating temperature range is 0°C to 40°C.
It uses a barrel plug with a 2.1mm I.D. x 5.5mm O.D. x 9.5mm size and positive center polarization.
Yes, it is approved to EN/IEC/UL60601-1 edition with isolation levels satisfying the 2 MOPP requirements.
It meets EISA2007, CEC Efficiency Level V, and EU (EC) No 278/2009 Phase II.