| Category | |
| Manufacturer Name | |
| Applications | |
| Approval Agency | |
| Cord Length | |
| Current | |
| Efficiency | |
| Form | |
| Input Connector | |
| Input Type | |
| Input Voltage | |
| Medical Grade | |
| No Load Power Consumption | |
| Operating Temperature | |
| Output Connector | |
| Output Power | |
| Overload Protection | |
| Polarization | |
| Power (Watts) | |
| Region Utilized | |
| Turn On Time | |
| Voltage |
The SL Power ME20A1203N01 is a medical-grade external power adapter from the ME20 series, delivering up to 20W of AC-DC power. This specific model provides a 12V output at 18W continuous power. It accepts a universal input voltage range of 90 to 264 VAC, making it suitable for international use.
The adapter features an IP22 rated enclosure, offering protection against solid objects and dripping water. It is designed for applications where stable power is critical, even in noisy or unstable AC mains environments. Key electrical specifications include a typical efficiency of 87% and a low no-load power consumption of less than 0.1W, meeting DoE Efficiency Level VI requirements.
Designed for reliability, the ME20A1203N01 boasts a Mean Time Between Failures (MTBF) of over 1,000,000 hours and an estimated E-Cap life of over 10 years. It includes comprehensive protection features such as overtemperature, overload, overvoltage, and short circuit protection, all operating in a hiccup mode with auto-recovery where applicable.
Mechanical specifications include a compact desktop form factor (Class II) with dimensions of 84.0mm x 47.0mm x 31.0mm. It utilizes an IEC 320-C8 input connector and a standard barrel plug output connector (2.5mm I.D. x 5.5mm O.D. x 9.5mm) with a positive center pin. The adapter complies with various safety and EMC standards, including IEC/EN/UL60601-1, EN55011, and FCC Part 15.109 Class B, and carries CE and UKCA marks.
The input voltage range is 100 to 240 VAC.
The output is 12V DC at 18W continuous.
The operating temperature is -20°C to 50°C.
It is certified to IEC/EN/UL60601-1 (3rd edition) and ANSI/AAMI ES60601-1, CSA CAN/CSA-C22.2 NO. 60601-1, and EN 60601-1 standards.
The efficiency level is VI (DoE Level VI).
Overload protection, overvoltage protection, short-circuit protection (all hiccup mode), and overtemperature protection (auto-recovery).
The dimensions are 84.0mm x 47.0mm x 31.0mm (3.31 x 1.85 x 1.22 inches).