The SL Power Electronics MC24-2.4-A is a linear AC DC converter designed for various applications, including medical use. It operates with a wide AC input range of 100 to 240 VAC and provides a stable 24V DC output at 2.4A, totaling 58W of power.
Key electrical specifications include a line regulation of ±0.05% and a load regulation of ±0.05%. The output ripple is a maximum of 3 mV peak-to-peak. Transient response is less than 50 microseconds for a 50% load change. The unit includes automatic current limit/foldback for short circuit protection and built-in overvoltage protection on 5V models.
This converter is chassis mounted and features adjustable output, remote sensing with built-in open sense lead protection (except HA series), and universal input. It operates within a temperature range of 0°C to 50°C at full load, derating linearly to 40% at 70°C. The transformer insulation meets Class F (155°C) standards.
Compliance includes CE, cURus, and TUV approvals, meeting UL1950, CSA C22.2 No. 950 L3, and EN60950 standards. It also meets medical standards UL2601-1, IEC60601-1, and CSA601.1, with medical leakage current not exceeding 10 µA. The unit exceeds FCC and CISPR22 Class B conducted emissions standards.
The input voltage accepts 100, 120, or 240 VAC with tolerance: +10%/-13% for 100/120/240 VAC and +12%/-11% for 215 VAC; tolerance for 230 VAC is +15%/-10%. Frequency is 47 to 63 Hz.
The output is 24 VDC at 2.4 A, adjustable ±5% minimum (except HA series). Power is 58 W.
Full rated operation from 0 to 50°C, derated linearly to 40% at 70°C.
The mounting type is chassis mount.
Short circuit protection via automatic current limit/foldback. Overvoltage protection is built-in on 5V models (set at 6.2 V ±0.4 V) and optional on other models.
Yes, remote sensing is provided on all models with open sense lead protection built-in (except HA series).
Commercial approvals: UL1950, CSA C22.2 No. 950 L3, TUV certified to EN60950. Medical approvals: UL2601-1, IEC60601-1, CSA601.1. Also CE, cURus, TUV marked.