| Category | |
| Manufacturer Name | |
| Applications | |
| Approval Agency | |
| Current | |
| Efficiency | |
| Features | |
| Height | |
| Input Frequency | |
| Input Voltage | |
| Length | |
| Medical Grade | |
| Mounting Type | |
| Number of Outputs | |
| Operating Temperature | |
| Output Power | |
| Power (Watts) | |
| Relative Humidity | |
| Standard Number | |
| Storage Temperature | |
| Type | |
| Voltage | |
| Width |
The EOS Power LFMWLT40-1004 is a 40 Watt Medical (MWLT) series AC/DC converter designed for universal input voltage ranging from 90 to 264 VAC across a frequency of 47 to 400 Hz. This open-frame converter offers a maximum output power of 40W and boasts an efficiency of 85% typical at both 120 VAC and 230 VAC input.
Key electrical specifications include line regulation of +/-0.3% and load regulation of +/-0.5% for V1. It features over-current protection at 130% typical and over-voltage protection for V1 only. The device operates within a temperature range of -20 to 70°C with derating and has a Mean Time Between Failures (MTBF) of 1.87 million hours.
Designed for medical applications, the LFMWLT40-1004 meets IEC60601-1 standards and offers isolation voltages of 4000 VAC (Input to Output), 1500 VAC (Input to Earth), and 500 VAC (Output to Earth). It is chassis mountable and cooled via convection. The dimensions are 4.00" L x 2.00" W x 1.12" H.
This model is suitable for various applications requiring reliable power conversion, with features like low leakage current (<250 µA) and low no-load power (<0.3 W). It also supports Class I & Class II options.
The input voltage range is 90-264 VAC.
The output voltage is 48 V and the output current is 0.83 A.
The operating temperature range is -20°C to 70°C with derating.
Dimensions are 101.6mm x 50.8mm x 28.5mm (4.00 x 2.00 x 1.12 inches) and it is designed for chassis mount.
It is certified to IEC 60601-1 with 2MOPP isolation between input and output.
Overcurrent protection at 130% typical, overvoltage protection at 130% typical for V1, and short circuit protection with autorecovery.
Efficiency is 85% typical and no-load power consumption is less than 0.3 W.